Sorry, you need to enable JavaScript to visit this website.

Study of SAW gas sensing versus gas concentration

TitoloStudy of SAW gas sensing versus gas concentration
Tipo di pubblicazionePresentazione a Congresso
Anno di Pubblicazione1999
AutoriAnisimkin, I.V., Anisimkin V.I., Caliendo C., Verona E., and Penza Michele
Conference NameProceedings of the IEEE Ultrasonics Symposium
EditoreIEEE, Piscataway, NJ, United States
Conference LocationCaesars Tahoe, NV, USA
Parole chiaveAcoustic surface wave devices, Calibration, Chemical sensors, Crystal orientation, Gas concentration, Lithium compounds, Piezoelectric single crystals, Polyvinyl alcohols, Single crystals, Sorbent membranes, Temperature coefficient of velocity, Velocity
Abstract

An analytical expression for the SAW sensor gas response versus gas concentration (calibration curve) is deduced, at low gas concentrations, on using the perturbation approach. The calibration curve, the sensitivity (its slope) and the detection limit (threshold value) for a given gas and sorbent membrane, can be controlled by a proper choice of the substrate material and its crystallographic orientation. We show how this property takes place from the anisotropy of the SAW propagation on piezoelectric single crystals, allowing to change the partial contributions to the SAW response through mechanical displacements, coupling constant and temperature coefficient of velocity (TCV). The theoretical predictions are experimentally verified on devices implemented on different orientations of quartz and LiNbO3 substrates, using Pd, Pd:Ni and polyvinyl alcohol (PVA) as sorbent films and humid air as analyte. The lowest threshold value of RH (0.01%) and the highest sensitivity (49 ppm/%) are detected in the PVA/yz-LiNbO3 structure, whose sensitivity is about two orders of magnitude higher than that observed in uncoated LiNbO3.

URLhttps://www.scopus.com/inward/record.uri?eid=2-s2.0-0033293696&partnerID=40&md5=37950e212ff5704d46d81f878ab1db08
Citation KeyAnisimkin1999485